This report has been identified as b.Braun medical inc.Internal report number (b)(4).The evaluation of the trend data record showed that about 2:29 [h:mm] after therapy start, the venous pressure decreased from about 190 mmhg to about 140 mmhg.Since the lower venous pressure limit was 138 mmhg, no alarm was triggered.This might have been the time when the venous access dislocated.At 2:31 [h:mm] the alarm "arterial pressure - lower limit" was triggered.The blood pump stopped and the dialysis machine switched into patient-safe mode.As a follow-up alarm due to the stop of the blood pump, the alarm "venous pressure - lower limit - check access" was triggered.The alarms were acknowledged by the operator.Twenty-three (23) minutes later the alarm "venous pressure - upper limit" was triggered and also acknowledged by the operator.After another eight (8) minutes the alarm "venous pressure - lower limit - check access" was triggered and acknowledged by the operator.This was likely the time the nurse detected the dislocation of the venous access.Until the trend data record ended abruptly, about 30 minutes later, numerous air and pressure alarms were triggered.The evaluation of the trend data record does not show any malfunction of the dialog+ dialysis machine.It operated as intended.The technical inspection of the dialog+ machine showed that the blood pressure measuring (abpm) module was inoperative, otherwise the dialog+ machine operated as intended.The therapy in question was performed without using the abpm module.Thus it did not have any influence on the situation described above.Furthermore, even if it would have been used, it would not have had an influence.As outlined in iec/tr 62653 (guideline for safe operation of medical equipment used for haemodialysis treatments), the venous pressure monitor of a dialysis machine may not reliably detect leaks, blood tubing separation from the blood access device, or needle dislodgement.Little or no pressure change may occur depending on the circumstance.It is unlikely but possible for a leak to occur, the blood tubing to separate, or access needle to dislodge without venous pressure alarm.For that reason, the patient's safety is only ensured by careful monitoring by the operator.Dependent on the venous pressure drop and the alarm limits set, the pressure monitoring system of the dialog+ machine as dialysis machines in general, might not reliably detect a disconnection or dislodgement of the venous patient access.Therefore, the instructions for use of the dialog+ dialysis machine shows the following warning: risk to patient due to blood loss if cannulas get disconnected or dislodged! standard monitoring function of the dialysis machine cannot ensure detection if the cannulas get disconnected or dislodged.Ensure that the access to the patient always remains fully visible during therapy.Ensure that cannulas are adequately fixed.Regularly check patient access.Venous lower limit should be set to greater than or equal to 20 mmhg in tsm (technical service mode).If additional pertinent information becomes available a follow-up report will be filed.
|