• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA; BTT Back to Search Results
Model Number OPT944
Device Problems Break (1069); Failure to Deliver (2338); Material Split, Cut or Torn (4008)
Patient Problem Insufficient Information (4580)
Event Date 09/22/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of retrieving further information and the subject opt944 optiflow + adult nasal cannula for evaluation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in (b)(6) reported that a patient was in a critical condition when the tubing of an opt944 optiflow + adult nasal cannula was damaged during use and stopped delivering oxygen.There were no further reported patient consequences.
 
Manufacturer Narrative
(b)(4).The opt944 optiflow + adult nasal cannula is an interface used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 optiflow + adult nasal cannula was returned to fisher & paykel healthcare (f&p) where is was visually inspected.Additional information with regards to the reported event was also requested from the hospital; however, no further information was provided.Results: visual inspection of the returned complaint opt944 optiflow + adult nasal cannula revealed that the tubing was pulled apart and stretched in several places.Conclusion: we are unable to determine the cause of the reported event; however, the damage observed was most likely caused by the tubing of the opt944 optiflow + adult nasal cannula being pulled.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is rejected.The subject opt944 optiflow + adult nasal cannula would have met the required specifications at the time of production.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "do not crush or stretch tube, to prevent loss of therapy." - "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event Description
A healthcare facility in texas reported that a patient was in a critical condition when the tubing of an opt944 optiflow + adult nasal cannula was damaged during use and stopped delivering oxygen.There were no further reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key12693490
MDR Text Key278335497
Report Number9611451-2021-01189
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2021
Initial Date FDA Received10/25/2021
Supplement Dates Manufacturer Received12/06/2021
Supplement Dates FDA Received12/15/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-