Model Number CLV-190 |
Device Problems
Fracture (1260); Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The evaluation confirmed the reported power switch malfunction as the power switch cannot stay on due to faulty power switch and also there is a crack on the front panel.Additionally, the lamp output is out of specification and there is corrosion inside the air tube.The investigation is ongoing; therefore, the root cause of the reported issue/malfunction cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
|
|
Event Description
|
The service center was informed by the biomedical equipment technician at the user facility that the evis exera iii xenon light source 's "power switch is damaged, wont stay on, crack on front panel" was observed during preparation for use.No patient involvement or harm was reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation the root cause of the faulty power switch could not be identified.Olympus will continue to monitor the field performance of this device.
|
|