RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2021
Event Type  malfunction  
Event Description
It was reported that the ventilators blower stalled.There was no patient involvement.The customer troubleshot with technical support and found in the device diagnostic logs error code blower stalled, then set the pressure to 1 and noted that the blower revolutions per minute (rpm) goes to 3000.The customer noted a brown residue in the blower.Upon a follow up the customer noted that the blower boots, internal hoses were discolored, and the gas outlet is clear.The customer was advised to replace the blower, tubing, and rubber boot kit.Further information is pending.
 
Manufacturer Narrative
Multiple attempts were made to obtain information regarding the repair and operational status with no response from the reporter and cannot be determined.If additional information is later obtained, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key12697951
MDR Text Key278382217
Report Number2031642-2021-05379
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/27/2021
Initial Date FDA Received10/26/2021
Supplement Dates Manufacturer Received11/17/2021
Supplement Dates FDA Received12/15/2021
Date Device Manufactured12/11/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-