RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the ventilators blower stalled.There was no patient involvement.The customer troubleshot with technical support and found in the device diagnostic logs error code blower stalled, then set the pressure to 1 and noted that the blower revolutions per minute (rpm) goes to 3000.The customer noted a brown residue in the blower.Upon a follow up the customer noted that the blower boots, internal hoses were discolored, and the gas outlet is clear.The customer was advised to replace the blower, tubing, and rubber boot kit.Further information is pending.
|
|
Manufacturer Narrative
|
Multiple attempts were made to obtain information regarding the repair and operational status with no response from the reporter and cannot be determined.If additional information is later obtained, a supplemental report will be submitted.
|
|