TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Model Number IPN913776
Device Problem Insufficient Heating (1287)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of the reported failure mode was conducted, and 8 devices were taken from current production (425-00 concha neptune lot # 73j210017n) at the manufacturing facility and were inspected.During the test, issue reported "unit is not heating" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
It was reported "neptune heater fails to warm up".No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key12712685
MDR Text Key278830767
Report Number3003898360-2021-00977
Device Sequence Number1
Product Code BTT
UDI-Device Identifier14026704646811
UDI-Public14026704646811
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial
Report Date 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberIPN913776
Device Catalogue Number425-00
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/06/2021
Initial Date FDA Received10/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-