• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2021
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital on (b)(6) 2021 for a follow-up.During interrogation of the pacemaker, it was noted that the radio frequency (rf) telemetry was not functioning.Since patient is dependent, software upgrade was not performed to resolve the issue.The pacemaker will be interrogated using inductive telemetry going forward.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12721423
MDR Text Key279097330
Report Number2017865-2021-35467
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberA000105345
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2021
Initial Date FDA Received10/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age83 YR
-
-