EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER; CATHETER, OXIMETER, FIBER-OPTIC
|
Back to Search Results |
|
Model Number 777F8 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
One 777f8 with monoject limited volume syringe was received for evaluation.The report of balloon ruptured was confirmed.The balloon was found to be ruptured and the ruptured edges did not appear to match up.All the through lumens were patent without any leakage or occlusion.There was no visible damage or inconsistency observed from the catheter body and syringe returned.A review of the manufacturing records indicated that the product met specifications upon release.As part of the manufacturing process, 100% balloon inflation and visual inspection is performed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.(b)(4).
|
|
Event Description
|
It was reported that during preparation, while testing this catheter, bubbles formed in the water and the balloon was ruptured.There was no patient involvement.
|
|