EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER; CATHETER, OXIMETER, FIBER-OPTIC Back to Search Results
Model Number 777F8
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One 777f8 with monoject limited volume syringe was received for evaluation.The report of balloon ruptured was confirmed.The balloon was found to be ruptured and the ruptured edges did not appear to match up.All the through lumens were patent without any leakage or occlusion.There was no visible damage or inconsistency observed from the catheter body and syringe returned.A review of the manufacturing records indicated that the product met specifications upon release.As part of the manufacturing process, 100% balloon inflation and visual inspection is performed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.(b)(4).
 
Event Description
It was reported that during preparation, while testing this catheter, bubbles formed in the water and the balloon was ruptured.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER
Type of Device
CATHETER, OXIMETER, FIBER-OPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key12721743
MDR Text Key282964665
Report Number2015691-2021-06068
Device Sequence Number1
Product Code DQE
UDI-Device Identifier00690103146554
UDI-Public(01)00690103146554(17)230423(11)210423(10)63737378
Combination Product (y/n)N
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/23/2023
Device Model Number777F8
Device Catalogue Number777F8
Device Lot Number63737378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2021
Initial Date FDA Received10/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-