This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Two (2) pictures are attached, during the analysis carried out, a top view of a product could be observed in the first image, then in the second image the inflated cuff can be observed.The returned sample was visually inspected at 12 to 16 inches and normal conditions of illumination.Per visual inspection, no damage was found on the components, no contamination, no failure mode conditions were observed, the sample meets all the specifications.No root cause could be determined since the complaint was not confirmed due sample provided does not shows failure mode reported, however current process mitigation related to failure mode reported were referred.No corrective actions are required since the complaint was not confirmed since sample successfully passed.
|