Model Number 101/870/075CZ
Device Problems Leak/Splash (1354); Gas/Air Leak (2946)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2021
Event Type  malfunction  
Event Description
It was reported that during the use of the product, leakage of air was observed.The customer suspected it was caused by the anomaly in the pilot balloon.No patient injury.
 
Manufacturer Narrative
Other, other text: h10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection did found the inflation line detached from pilot balloon was observed and a lack of solvent was detected; thus, the failure mode reported is confirmed.The occurrence of this failure condition could be caused by -not enough thf solvent at joint-.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
TRACHEOSTOMY TUBE
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
david halverson
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key12725664
MDR Text Key279261319
Report Number3012307300-2021-10496
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076219
UDI-Public15019517076219
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/870/075CZ
Device Catalogue Number101/870/075CZ
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/08/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/07/2021
Initial Date FDA Received10/31/2021
Supplement Dates Manufacturer Received01/17/2022
Supplement Dates FDA Received01/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-