The reported event was confirmed manufacturing related.1 sample was confirmed to exhibit the reported failure.The device has not met specifications.The product was used for treatment.The product had caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used temperature sensing silicone foley catheter.Visual inspection of the sample noted no obvious visible observations.Attempted to flush the catheter with methylene blue solution, and solution would not advance through the funnel due to lumen blockage.This is out of specification per inspection procedure, which states, "check for defects: excessive material (over fill), bubbles/voids, splits and/or blemishes".A potential root cause for this failure could be tooling misalignment.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: a labeling review was not performed because labeling could not have prevented the reported failure.The actual/suspected device was inspected.
|
It was reported that the urine did not flow after the foley catheter was placed during surgery.The customer requested to investigate if there was something stuck in it.Per follow up information received via ibc on 19oct2021, there any blockage in the catheter was unknown and no, they could obtain no additional information of the issue.
|