STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR W/ SIDNE (ASIA); INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Catalog Number 0620040614
Device Problems Suction Problem (2170); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2021
Event Type  malfunction  
Event Description
It was reported that there was loss of insufflation during procedure.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that there was loss of insufflation during procedure.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: loss of insufflation.Probable root cause: pressure sensor malfunction / out of calibration software malfunction use error system design unwanted movement of internal components / wiring power button inadvertently turned off tubeset/gas supply inadvertently detached/loose loss of power pressure button does not disengage electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge power supply malfunction flow sensor malfunction leaks from internal connections or seals ppv failure manufacturing/ service error the reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE XL HIGH FLOW INSUFFLATOR W/ SIDNE (ASIA)
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12733709
MDR Text Key281024583
Report Number0002936485-2021-00589
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K063367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0620040614
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/08/2021
Initial Date FDA Received11/01/2021
Supplement Dates Manufacturer Received10/08/2021
Supplement Dates FDA Received12/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-