Model Number 24674
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2021
Event Type  malfunction  
Manufacturer Narrative
Age at time of event - 18 years or older.A mustang device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 8mm proximal from the proximal markerband.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.
 
Event Description
Reportable based on device analysis completed on 15oct2021.It was reported that balloon leak occurred.Vascular access was obtained via the femoral artery.The 10x2cm target lesion was located in the iliac vein.A 10.0 x 40, 135cm mustang balloon catheter was advanced for dilation.During inflation, a leak was noticed in the distal part of the balloon.The device was removed and the procedure was completed with another of the same device.There were no patient complications reported and the patient was stable.However, returned device analysis revealed balloon pinhole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12737193
MDR Text Key279579896
Report Number2134265-2021-13332
Device Sequence Number1
Product Code FGE
UDI-Device Identifier08714729793823
UDI-Public08714729793823
Combination Product (y/n)N
Reporter Country CodeUY
PMA/PMN Number
K141521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/22/2024
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0027192660
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/15/2021
Initial Date FDA Received11/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-