C.R. BARD, INC. (BASD) -3006260740 UNKNOWN POWER PORT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number UNKNOWN POWER PORT
Device Problems No Display/Image (1183); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history record could not be performed.The return of the sample is pending.However, images were provided for review.The investigation of the reported event is currently underway.
 
Event Description
It was reported that after a port placement procedure, the port was allegedly not visualized under imaging.There was no reported patient injury.
 
Event Description
It was reported that after a port placement procedure, the ct marker was allegedly not visualized under imaging.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the sample was not returned for evaluation.Three medical images were provided for review.The x-ray images shows a port likely along the chest.Image 2 appears to show a ct marker in the middle of the port.Additionally, there are three marks around the port signifying the outer aspects of the device and these are visualized in each of the images.However it is unknown whether they could not visualize the outer or inner markers of the device.Therefore, the investigation is inconclusive for the reported poor visualization issue as the exact circumstances at the time of the reported event are unknown.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN POWER PORT
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12742045
MDR Text Key279814606
Report Number3006260740-2021-04636
Device Sequence Number1
Product Code LJT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN POWER PORT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/07/2021
Initial Date FDA Received11/03/2021
Supplement Dates Manufacturer Received12/04/2021
Supplement Dates FDA Received12/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial