MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M Back to Search Results
Model Number BI70000027230
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2021
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Other relevant device(s) are: product id: bi31000160, lot/serial #: unknown.The manufacturer representative went to the site to test the imaging system.After they adjusted the lever on the lower left site from the front cover there was no issue.The adjustment was performed by the site while on the phone with the manufacturer representative.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the gantry wagged to the left and right and that the system moved during the scan.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
MDR Report Key12745219
MDR Text Key279877400
Report Number3004785967-2021-01250
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
PMA/PMN Number
K173664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000027230
Device Catalogue NumberBI70000027230
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/18/2021
Initial Date FDA Received11/03/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-