Model Number 0998-00-0800-53
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate this unit and was able to duplicate the reported issue of not "slaving signal to monitor" or not sending the signal to a secondary monitor.To correct the problem, the fse replaced the front end board.The unit passed all functional and safety tests according to factory specifications.The iabp was returned to the customer for clinical use.Upon completion of our investigation, a supplemental report will be submitted.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) will not send the signal to a secondary monitor (slave to monitor).There was no harm or injury to the patient and no adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, b6, b7, g4, g7, g8, h2, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: g1 (contact person ¿ mfg site).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12746349
MDR Text Key279922094
Report Number2249723-2021-02542
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/15/2021
Initial Date FDA Received11/03/2021
Supplement Dates Manufacturer Received09/06/2022
Supplement Dates FDA Received09/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-