A getinge field service engineer (fse) was dispatched to evaluate this unit and was able to duplicate the reported issue of not "slaving signal to monitor" or not sending the signal to a secondary monitor.To correct the problem, the fse replaced the front end board.The unit passed all functional and safety tests according to factory specifications.The iabp was returned to the customer for clinical use.Upon completion of our investigation, a supplemental report will be submitted.
|
Updated fields: b4, b6, b7, g4, g7, g8, h2, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: g1 (contact person ¿ mfg site).
|