• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the implantable cardiac monitor (icm) had reached end of service (eos) prior to being implanted.The device remains in use. no patient complications have been reported as a result of this event.
 
Event Description
It was further reported that the implantable cardiac monitor (icm) is no longer in use.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated lock-up in an integrated circuit.Hybrid analysis concluded that the end of service (eos) condition was due to a failing analog to digital converter (adc) state machine locking up in the l466 ic of the stacked chip scale package (scsp).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12750505
MDR Text Key280071018
Report Number9614453-2021-04181
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169845749
UDI-Public00643169845749
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/14/2022
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Initial Date Manufacturer Received 11/01/2021
Initial Date FDA Received11/04/2021
Supplement Dates Manufacturer Received12/20/2021
02/01/2022
Supplement Dates FDA Received01/06/2022
02/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight91 KG
-
-