All available information was investigated, and the returned device analysis confirmed the reported difficult to open or close the clip.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.All available information was investigated and returned device analysis, a cause for the reported inability to open the clip could not be determined.The observed twisted on the l-lock tabs, scratched on the l-lock tabs, cracked on the l-lock tabs, and frayed on the lock line appear to be cascading effect of the troubleshooting.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
This is being filed to report the cracked l-lock tabs and frayed lock line noted on the returned device.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 3+.The clip delivery system (cds) was advanced into the patient and the clip arms tested however they would not open.Troubleshooting was performed however the clip arms would not open therefore the cds was removed.Another clip was implanted, reducing mr to 1.There was no clinically significant delay in the procedure and no adverse patient effects.Return device analysis noted the l-lock tabs were observed to be scratched, twisted, and cracked.The lock line was noticed to be frayed.No additional information was provided.
|