Catalog Number 102451300 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Ossification (1428); Unspecified Musculoskeletal problem (4535)
|
Event Date 10/21/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Revision of a sigma uni tkr.Reason for revision was due to impingement due to excess bone on the medial side of the tibial plateau.No issue with the implants themselves.The poly was removed in order to obtain access so the bone could be cleared.A new poly of the same thickness was then implanted.
|
|
Event Description
|
Additional information received indicated that there was no surgical delay.The affected side was the left side.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Udi : (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed, and there is no evidence that the complaint is product related depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|