BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 87035 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported that during an ablation procedure involving an intellamap orion high resolution mapping catheter to treat atrial fibrillation there was no irrigation flow.No error messages occurred.The catheter from the body and confirmed that the irrigation was not flowing from the tip.They attempted to send saline through the catheter again and found that the catheter lumen was blocked.The catheter was exchanged and the procedure was completed without patient complications.The catheter has been received for analysis by boston scientific.
|
|