• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TAEWOONG MEDICAL CO.,LTD. NITI-S PYLORIC & DUODENAL COVERED STENT; PYLORIC & DUODENAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TAEWOONG MEDICAL CO.,LTD. NITI-S PYLORIC & DUODENAL COVERED STENT; PYLORIC & DUODENAL STENT Back to Search Results
Model Number DCT2012BP
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problems Vomiting (2144); Abdominal Distention (2601)
Event Type  Injury  
Manufacturer Narrative
It was reported that after 17 days of stent placement with clipping with another device, the stent was found being migrated.It is hard to review suspected device's dhr, because the serial no.Was not checked.Migration can occur in any company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.It is difficult to identify the exact root cause because the device was not returned, and it is difficult to reconstruct the situation at the time of procedure.However, based on the description "stent was found being migrated" it is assumed stent migration occurred due to severe pressure at the patient's lesion, peristalsis of organs, foreign substances such as food, clipping with another device and other factors complexly, and it is considered this caused the patient to vomit and have feeling of abdominal pressure.After that, it is considered the stent was removed.Through the user manual by taewoong, it is stated that "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: stent migration".This suspected device is not registered in the us but we will continuously monitor the same or similar customer complaints through accurate analyses.
 
Event Description
Dct2012bp was placed in the duodenal bulb and otsc system was clipped to fix the stent to prevent the migration.After 17 days of the stent placement, the patient complained of vomit and feeling of abdominal pressure, so the patient stayed in the hospital.When checking, the stent was found being migrated, so it was removed by the small intestinal scope, and additional stent placement was performed to finish the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S PYLORIC & DUODENAL COVERED STENT
Type of Device
PYLORIC & DUODENAL STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
lee
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
MDR Report Key12757129
MDR Text Key281742924
Report Number3003902943-2021-00042
Device Sequence Number1
Product Code MUM
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberDCT2012BP
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/11/2021
Initial Date FDA Received11/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-