Model Number GF-UCT180
Device Problems Break (1069); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the preparing for use of the subject device, it was found that the ultrasound transducer of the subject device was physically damaged and a part of the pink coating peeled off on the ultrasound probe unit.The occurrence date of the event is unknown, and there was no report of patient injury associated with this event.
 
Manufacturer Narrative
The subject device was not returned to omsc for evaluation but was returned to olympus medical systems (b)(4).(b)(4) checked the subject device and found that the reported phenomenon was duplicated due to physical stress by dropping and/or knocking the affected part to a hard surface/object and applying a chemical stress during the cleaning/sanitization process.Also, during the evaluation, (b)(4) found that there was a foreign material on groove or gap of the distal end due to insufficient cleaning.The exact cause has been under investigation.Therefore, the exact cause of the reported phenomena could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12757584
MDR Text Key281834793
Report Number8010047-2021-14131
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170341793
UDI-Public04953170341793
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/08/2021
Initial Date FDA Received11/07/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-