Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Great Vessel Perforation (2152); Cardiac Perforation (2513)
Event Date 10/29/2021
Event Type  Injury  
Event Description
During transseptal puncture, the brk needle and guidewire punctured through the aorta; the dilator and the sheath remained in the right atria.An echocardiogram showed no tamponade or effusion in the pericardial space.
 
Event Description
During an atrial fibrillation ablation procedure, the brk needle and guidewire punctured through the aorta; the dilator and the sheath remained in the right atria.An echo was done just after the perforation and twenty minutes later; no effusion was present.A scan was performed one hour post procedure with no effusion or tamponade noted.There were no performance issues with any abbott devices.
 
Manufacturer Narrative
Additional information: g3, h2, h3 the results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported perforation remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSSEPTAL NEEDLE, BRK SERIES
Type of Device
TRANSSEPTAL NEEDLE
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12758263
MDR Text Key280448103
Report Number3008452825-2021-00517
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/29/2021
Initial Date FDA Received11/07/2021
Supplement Dates Manufacturer Received11/24/2021
Supplement Dates FDA Received12/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-