• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ULTRA SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINE ULTRA SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/070CZ
Device Problem Suction Problem (2170)
Patient Problem Insufficient Information (4580)
Event Date 10/01/2021
Event Type  malfunction  
Event Description
It was reported that during the use of the product, the customer noticed that no suction was able to be performed.So he changed the product to another one.He then checked the removed product for occlusion by injecting air into it using a syringe and a burst sound was heard and he confirmed air was passing through the removed product.The customer suspected there was an occlusion in the suction line.No patient injury.
 
Manufacturer Narrative
(b)(4) unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.Therefore, according pfmea the occurrence of this failure condition could be caused by: thf (tetrahydrofuran) excess, due to tube submerged more than one time into the solvent dispenser.An awareness complaint notification to the production personnel was conducted by quality engineer on 26-jan-2022 for the defect reported by the customer.Dhr review was done, no issues related to the original complaint were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ULTRA SUCTIONAID
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
akasaka, minato-ku, tokyo
minneapolis, MN 55442
MDR Report Key12764064
MDR Text Key280469484
Report Number3012307300-2021-10817
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076134
UDI-Public15019517076134
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/070CZ
Device Catalogue Number101/860/070CZ
Device Lot Number4052498
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/09/2021
Initial Date FDA Received11/07/2021
Supplement Dates Manufacturer Received02/08/2022
Supplement Dates FDA Received03/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-