MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE Back to Search Results
Model Number EB15-J10
Device Problems Obstruction of Flow (2423); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2021
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.
 
Event Description
A brush stuck inside the inlet and the op-channel.This happened during the pre cleaning.No patient is involved / hurt.The time of event is reprocessing.There was no report of patient harm.
 
Manufacturer Narrative
Evaluation summary: we checked the returned unit and confirmed that the biopsy inlet t-piece clogged.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility.In addition, we confirmed that the segment compressed; however, it is not the main cause, and/or irrelevant to the alleged complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO BRONCHOSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12764321
MDR Text Key284975859
Report Number9610877-2021-01450
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K200678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEB15-J10
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/27/2021
Initial Date FDA Received11/07/2021
Supplement Dates Manufacturer Received10/27/2021
Supplement Dates FDA Received12/06/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-