CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
|
Back to Search Results |
|
Catalog Number 930480NSB |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the liquid in the prep is crystalized and cannot be used.Verbatim: i work for (b)(6) and we purchase your chloraprep app 1ml item code #930480nsb.I wanted to notify you that we have received a complaint from one of our customers.The complaint report stated the following: several nurses have recently reported an issue with the chloraprep swab inside of the iv start kits.The problem is that the liquid in the prep is crystalized and cannot be used.When it is popped, the liquid is in several pieces.A sample has not been received at this time.Have you received similar complaints for this issue? if you have additional insight or information that could be used for the investigation, please let me know.
|
|
Manufacturer Narrative
|
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
|
|
Manufacturer Narrative
|
A photo was provided for evaluation.Unfortunately, the failure mode could not be verified with the photo provided.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of the glass, it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excess handling.The production record review was completed for batch/lot 1054122 and no non-conformities, failures, deviations, or rework activities occurred during the manufacturing of this lot similar to the reported issue or that could have contributed to the reported failure mode.Records reviewed indicate that the lot passed all the in-process inspections.No further actions are required.This failure will continue to be tracked and trended.H3 other text : see narrative below.
|
|
Event Description
|
It was reported that the liquid in the prep is crystalized and cannot be used.Verbatim: per email 14-oct-2021 how many devices experienced the reported issue? 1 ea was reported are any of the affected devices or photos displaying the reported issue available? the photograph below is the only photograph that was provided.What is the address of the sample¿s location? a sample has not been received at this time.Was there any evidence of the ampoule broken prior to handling? could glass shards be seen when held up to a light or heard when shaken? it was noted in the complaint that when it is popped, the liquid is in several pieces.I can ask to see if there is additional information about glass shards/ampoule breaking prior to handling.I work for division 20 (centurion medical products) at medline and we purchase your chloraprep app 1ml item code #930480nsb.I wanted to notify you that we have received a complaint from one of our customers.The complaint report stated the following: several nurses have recently reported an issue with the chloraprep swab inside of the iv start kits.The problem is that the liquid in the prep is crystalized and cannot be used.When it is popped, the liquid is in several pieces.The photograph below was provided.However, a sample has not been received at this time.Have you received similar complaints for this issue? if you have additional insight or information that could be used for the investigation, please let me know.
|
|