• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ALLIANCE II; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ALLIANCE II; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number M00550601
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an alliance inflation syringe was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, it was noted that the pressure table needle did not move, and the specific pressure could not be seen.The procedure was completed with another alliance inflation syringe.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Block e1: this event was reported by the distributor.The physician is: dr.(b)(6).Block h6: device code a0902 captures the reportable event of gauge reading inaccurate.Block h10: investigation results a visual examination of the returned complaint device found that the gauge needle was at 0.5 atm when received, which confirms the reported event.The device was received with its extension tube.Functional evaluation was performed, and the device was pressurized to 10 atm using 35 ml of sterile water; no leak was observed.The device was able to hold the pressure but gauge was not indicating an accurate value.Based on the available information, it's probable that during the unpackaging or during the procedure the gauge received any kind of hit which could affect the integrity of the device.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event Description
It was reported to boston scientific corporation that an alliance inflation syringe was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, it was noted that the pressure table needle did not move, and the specific pressure could not be seen.The procedure was completed with another alliance inflation syringe.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLIANCE II
Type of Device
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12775631
MDR Text Key284982007
Report Number3005099803-2021-05776
Device Sequence Number1
Product Code MAV
UDI-Device Identifier08714729129332
UDI-Public08714729129332
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K922573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2023
Device Model NumberM00550601
Device Catalogue Number5060-05
Device Lot Number0027074761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/21/2021
Initial Date FDA Received11/08/2021
Supplement Dates Manufacturer Received11/26/2021
Supplement Dates FDA Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age36 YR
Patient SexMale
Patient Weight69 KG
-
-