Model Number M00550601 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an alliance inflation syringe was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, it was noted that the pressure table needle did not move, and the specific pressure could not be seen.The procedure was completed with another alliance inflation syringe.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Block e1: this event was reported by the distributor.The physician is: dr.(b)(6).Block h6: device code a0902 captures the reportable event of gauge reading inaccurate.Block h10: investigation results a visual examination of the returned complaint device found that the gauge needle was at 0.5 atm when received, which confirms the reported event.The device was received with its extension tube.Functional evaluation was performed, and the device was pressurized to 10 atm using 35 ml of sterile water; no leak was observed.The device was able to hold the pressure but gauge was not indicating an accurate value.Based on the available information, it's probable that during the unpackaging or during the procedure the gauge received any kind of hit which could affect the integrity of the device.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
|
Event Description
|
It was reported to boston scientific corporation that an alliance inflation syringe was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, it was noted that the pressure table needle did not move, and the specific pressure could not be seen.The procedure was completed with another alliance inflation syringe.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
|