A patient fell from the mobile x-ray system table (type prognost xs) onto the bucky holder of the x-ray u-arm system used underneath.This is because the table top tilted when the patient sat on the table to take off his shoes.The table top bent towards the bucky holder underneath and the table stopped.Due to the low height of the fall, the patient was not injured in the incident and was able to leave the examination room independently and without any impairments.Third persons were also not injured.Besides the damage to the table, only damage to the x-ray u-arm system was reported to the manufacturer.After informing the manufacturer, the x-ray system table was immediately ordered to be taken out of service.This was implemented by the health facility; the affected table was replaced on (b)(6) 2021.
|