MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Contamination (1120)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that a thread like foreign material was found on an iol after implantation.The surgery was completed without product replacement.The foreign material remains in the eye.There are no plans to remove it with reoperation.The cause and effect is unknown.Additional information has been requested.
 
Manufacturer Narrative
The company cartridges were not returned for evaluation.A qualified handpiece and viscoelastic were indicated.It cannot be verified if the iol (intraocular lens) was in the qualified diopter range as the lot/serial number were not provided.The product investigation could not identify a root cause for the reported complaint.The product was not returned for evaluation.No determination can be made without physical evaluation of the complaint sample.It cannot be verified if the iol was in the qualified diopter range as the lot/serial number were not provided.Per the ifu (instructions for use): the company iol delivery system is for implantation of qualified company foldable iols.No unqualified lenses should be used with the company iol delivery system.The company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12780372
MDR Text Key280537096
Report Number1119421-2021-02088
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777639
UDI-Public00380659777639
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2021
Initial Date FDA Received11/09/2021
Supplement Dates Manufacturer Received01/05/2022
Supplement Dates FDA Received02/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACRYSOF IQ NATURAL SINGLEPIECE IOL.; IOL, UNSPECIFIED.; MONARCH III IOL DELIVERY SYSTEM, INJECTOR.; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE.; VISCOAT OPHTHALMIC VISCOSURGICAL DEVICE.
-
-