The reported event was confirmed manufacturing related.1 sample confirmed to exhibit the reported failure.The device had not met specifications.The product was used for patient treatment or diagnosis.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley.Visual inspection of the sample noted no obvious visible defects.The drainage lumen was flushed with methylene blue solution (3 drops 1% aq methylene blue per 100ml of distilled water) and did not advance past the funnel.The funnel seemed to be blocked with silicone.This does not meet the specification "check for defects: excessive material (over fill), bubbles/voids, splits and/or blemishes." a potential root cause for this failure could be ¿tooling wear¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The labeling review was not performed because labelling could not have prevented the reported failure.Correction: d,h.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|