• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION Vercise Genus¿; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION Vercise Genus¿; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-1216
Device Problems Device Displays Incorrect Message (2591); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/30/2021
Event Type  malfunction  
Event Description
The returned implantable pulse generator (ipg) was analyzed and r&d investigation found that the log records were not properly written due to an interruption of power to the pcba (printed circuit board assembly).Before the battery is attached, the pcba receives power from the ate (automated test equipment) test fixture through an external connector.Since log record writes by firmware are asynchronous to any ate test procedures, it is possible that power is removed from the pcba during a log record write.Any corrupt log records due to this reason should not indicate any issue with the stimulator.With all the available information, boston scientific concludes that the patients ipg having an error message was caused by the log records not being properly written due to an interruption of power to the pcba.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Vercise Genus¿
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key12797587
MDR Text Key280810379
Report Number3006630150-2021-06382
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729985044
UDI-Public08714729985044
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/19/2023
Device Model NumberDB-1216
Device Catalogue NumberDB-1216
Device Lot Number507324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/13/2021
Initial Date FDA Received11/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-