• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(M)
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2021
Event Type  malfunction  
Manufacturer Narrative
Cardioquip has not received the device for investigation.Once the investigation has been completed, cardioquip will file supplementary reports.
 
Event Description
Customer reports that the device is overheating while in use.When the customer biomed inspected the device, they found that the unit's cooling fan was not working as intended and that the device overheated for them as well shortly after circulating water.The device was swapped out for a replacement and the case continued as expected, the customer reports that this issue did not negatively impact patient health.
 
Manufacturer Narrative
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.The customer notified cardioquip, that a patient was involved during the malfunction although no patient information was communicated.Cardioquip's investigation determined that the heater, pump, tank, and heater chamber to be nonfunctional.Following replacement, the device passed inspection, and is fully operational.
 
Event Description
Customer reports device is overheating while in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key12798574
MDR Text Key281937953
Report Number3007899424-2021-00043
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial,Followup
Report Date 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCH-1000(M)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/12/2021
Initial Date FDA Received11/11/2021
Supplement Dates Manufacturer Received10/12/2021
Supplement Dates FDA Received10/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-