Catalog Number MG-STN3
Device Problem Malposition of Device (2616)
Patient Problems Failure of Implant (1924); Organ Dehiscence (2502)
Event Type  Injury  
Event Description
When the guide was removed after a guided surgery, it turned out that the implant in regio 18 was placed to much to buccal side, the bone was fenestrated.The implant had to be removed.
 
Manufacturer Narrative
Because a serious injury resulted, this event is reportable per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
The device was evaluated and found to be within specification.A dhr review was conducted with no discrepancies noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MGUIDE
Type of Device
ACCESSORIES, IMPLANT, DENTAL, ENDOSSEOUS
Manufacturer (Section D)
DENTSPLY IMPLANTS N.V.
research campus 10
hasselt limburg, B-350 0
BE  B-3500
Manufacturer (Section G)
DENTSPLY IMPLANTS N.V.
research campus 10
hasselt limburg, B-350 0
BE   B-3500
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key12803715
MDR Text Key280699426
Report Number3007362683-2021-00010
Device Sequence Number1
Product Code NDP
Combination Product (y/n)Y
Reporter Country CodeAU
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMG-STN3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/23/2021
Initial Date FDA Received11/12/2021
Supplement Dates Manufacturer Received10/23/2021
Supplement Dates FDA Received01/21/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-