• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRCRAFT MEDICAL LIMITED MCGRATH MAC VIDEO LARYNGOSCOPE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRCRAFT MEDICAL LIMITED MCGRATH MAC VIDEO LARYNGOSCOPE; LARYNGOSCOPE, RIGID Back to Search Results
Device Problem Power Problem (3010)
Patient Problem Airway Obstruction (1699)
Event Date 11/01/2021
Event Type  Injury  
Event Description
During a cardiac resuscitation on a patient that was apneic, an orotracheal intubation was attempted via use of the mcgrath mac video laryngoscope three times.During all three attempts, the scope powered off losing the video capabilities as well as the light on the blade thus causing the intubation attempt to secure an airway to fail.The serial number of said mcgrath mac video laryngoscope is (b)(4).The patient's airway was ultimately secured via a bls method.The blade would power off the video screen as well as the light.This power off occurred 5 times out of 13 intubation attempts during the in house testing.Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCGRATH MAC VIDEO LARYNGOSCOPE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
AIRCRAFT MEDICAL LIMITED
MDR Report Key12804243
MDR Text Key280822251
Report NumberMW5105279
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age57 YR
Patient SexMale
Patient Weight91 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-