Device Problem Use of Device Problem (1670)
Patient Problem Burn(s) (1757)
Event Date 10/18/2021
Event Type  malfunction  
Event Description
The user facility reported that an employee obtained a burn on their fingers while handling items that were processed in a v-pro 1 plus sterilizer.Medical treatment was sought and administered.
 
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the v-pro 1 plus sterilizer and found the unit to be operating properly.No repairs were required, and the unit was returned to service.The technician was informed that the employee subject of the event was not wearing proper ppe specifically gloves as stated in the operator manual.The v-pro 1 plus sterilizer operator manual states (1-2), "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." the technician counseled user facility personnel on the proper use and operation of the v-pro 1 plus sterilizer, specifically wearing proper ppe.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO 1 PLUS STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key12804910
MDR Text Key285354532
Report Number3005899764-2021-00058
Device Sequence Number1
Product Code MLR
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2021
Initial Date FDA Received11/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-