A steris service technician arrived onsite following the reported event to inspect the v-pro 1 plus sterilizer and found the unit to be operating properly.No repairs were required, and the unit was returned to service.The technician was informed that the employee subject of the event was not wearing proper ppe specifically gloves as stated in the operator manual.The v-pro 1 plus sterilizer operator manual states (1-2), "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." the technician counseled user facility personnel on the proper use and operation of the v-pro 1 plus sterilizer, specifically wearing proper ppe.No additional issues have been reported.
|