Model Number 101/860/075CZ
Device Problem Leak/Splash (1354)
Patient Problem Insufficient Information (4580)
Event Date 10/16/2021
Event Type  malfunction  
Event Description
It was reported that after days after inserting the product into the patient, the customer noticed a spontaneous removal of it occurred.Also, air in the cuff was easily deflated.The customer then checked the removed product by immersing it in water, but no leakage of air was detected.No patient injury.
 
Manufacturer Narrative
Returned device was received for evaluation.During the evaluation of the device the customer reported condition was confirmed.Problem source is unknown.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
TRACHEOSTOMY TUBE
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
david halverson
6000 nathan lane north
akasaka, minato-ku, tokyo
minneapolis, MN 55442
MDR Report Key12808167
MDR Text Key280770528
Report Number3012307300-2021-11171
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076141
UDI-Public15019517076141
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/075CZ
Device Catalogue Number101/860/075CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/09/2021
Initial Date FDA Received11/13/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-