Model Number 10094137
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2021
Event Type  malfunction  
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee ceiling system.During an interventional procedure, the user reported that the system went down and into bypass.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation showed that the system behavior was due to a hardware defect.The image acquisition system (ias) was unable to operate due to a defective ias computer.After a reboot, the system switched to the bypass fluoro mode, as intended for such a case.This special mode, which is a mitigation of the problem, allows the system to continue generating imaging (fluoro).However, neither acquisition nor storage or further processing of data is possible.The image quality is reduced.Such a defect can only be eliminated by replacing the affected hardware.The complete ias pc was replaced by the service organization.Because the customer's affected pc was an older model (ias bb1), it was exchanged for a compatible model of a newer type (ias bb2).Following the replacement, the system operates as specified.The occurrence rate of the error pattern and the spare parts consumption of the affected part were checked.A possible error accumulation or even a systematic error that would lead to a corrective action of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE CEILING
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
6104486461
MDR Report Key12819077
MDR Text Key281286250
Report Number3004977335-2021-02986
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010052
UDI-Public04056869010052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10094137
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/28/2021
Initial Date FDA Received11/16/2021
Supplement Dates Manufacturer Received12/09/2021
Supplement Dates FDA Received12/09/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-