Model Number 10-88-60
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).Analysis of the pump serial read-out has been conducted.A motor control failure error was found stored in the microcontroller.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The processor board of the pump control panel was replaced as well as internal cables.However, the issue persisted thus, the motor control board and the motor power amplifier board were replaced to solve the problem.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 mast roller pump have a motor control failure error message during maintenance.There is no known patient involvement.
 
Manufacturer Narrative
H.10: through follow-up communication livanova learned that this event is a duplicate of another event already reported under reference 9611109-2021-00585 due to a mistake of the customer providing the wrong pump serial number for event reported under ref.9611109-2021-00585.As a consequence, this case has been voided and all relevant information have been provided under the above mentioned reference.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich, germany
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key12827578
MDR Text Key280939571
Report Number9611109-2021-00661
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K062396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-88-60
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/19/2021
Initial Date FDA Received11/17/2021
Supplement Dates Manufacturer Received11/22/2021
Supplement Dates FDA Received12/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-