The cori drill long attachment was returned for evaluation. a relationship between the reported event and the device was established. the reported event was visually and functionally confirmed. the long attachment was out of specifications.The torque was found to be on 8 oz.In, instead of 20 oz.In as design.Even when the reported issue was confirmed visually, a functional evaluation was performed.The long attachment got stuck in the drill.Once the torque was adjust, it operated as design.A supplemental mdr with the results of investigation will be sent.
|
Section h3, h6: the cori drill long attachment, p/n rob10015, sn (b)(6), intended for use in treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported event was visually and functionally confirmed.The long attachment was out of specifications.The torque was found to be on 8 oz.In, instead of 20 oz.Lb as design.Even when the reported issue was confirmed visually, a functional evaluation was performed.The long attachment got stuck in the drill.Once the torque was adjust, it operated as design.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is normal wear.Vibrations can caused the long attachment torque came out of specifications.A historical capa, hhe/pra, field action review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and that the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
|