• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-260V
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to (b)(4) but has not been returned to omsc.(b)(4) checked the subject device for evaluation.It could be confirmed that the inside of the light guide lens of the subject device was dirty.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at (b)(4) it was found the dirt inside the light guide lens of the subject device.The subject device had been returned to (b)(4) for repair of the leakage at the control body.The occurrence date of the event is unknown.There was no patient injury associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.[conclusion] it could not identify the root cause of the reported phenomenon.[consideration] based on the following information, it was presumed that the cause of the reported phenomenon was that dirt had entered because a gap was created in the adhesive part of the light guide lens due to physical stress or the like.-according to the inspection results of olympus korea, it was confirmed that the light guide lens was dirty.-according to ifu, caution is stated not to apply physical stress.-according to a similar former case, it had been presumed that the cause was the physical stress caused a gap in the adhesive part of the light guide lens and dirt got in.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12833429
MDR Text Key280961287
Report Number8010047-2021-14724
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberTJF-260V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/21/2021
Initial Date FDA Received11/17/2021
Supplement Dates Manufacturer Received12/13/2021
Supplement Dates FDA Received01/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-