• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/21/2021
Event Type  Injury  
Manufacturer Narrative
Implant date is estimated to have occurred in 2019.Further information was requested but not received.
 
Event Description
The patient was admitted to the emergency room on (b)(6) 2021 after an attempted suicide and gunshot wound to the implanted pacemaker.The pacemaker went into back-up vvi (bvvi) mode and the patient felt phrenic nerve stimulation.Then, on (b)(6) 2021, the device was explanted after it was unable to be interrogated.It is unknown if the pacemaker could not be interrogated due to gunshot wound or not.Further information was requested but not received.
 
Manufacturer Narrative
The reported events of no pacing and failure to interrogate were confirmed.The device was received from the field with no output, no telemetry, and battery having reached end of service.Field information indicates the pacer tripped to backup mode due to a gunshot triggered by the patient causing a large dent on the device.The device image could not be saved due to lack of telemetry and data was requested to the field, but it could not be obtained.Analysis performed after cutting open the device case and replacing the power source indicated high current drained by the hybrid circuitry.Further analysis isolated the high current to a component on the hybrid circuitry, which depleted the battery prematurely, consistent with the reported field events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12843921
MDR Text Key281013115
Report Number2017865-2021-37182
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000066238
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/12/2021
Initial Date FDA Received11/19/2021
Supplement Dates Manufacturer Received02/28/2022
Supplement Dates FDA Received02/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-