MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problems Entrapment of Device (1212); Material Frayed (1262); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Event Description
During a focal atrial tachycardia procedure, when attempting to remove the mapping catheter from the patient, it became entangled with another catheter.When removing the mapping catheter, it descended the inferior vena cava until it reached the introducer, but as the other catheter was still placed in between the splines of the mapping catheter, it was not possible to remove the catheter from the patient.Because of such, one of the splines became frayed and an electrode displaced.The catheters were eventually separated and the mapping catheter was removed.The second catheter still functioned properly and was not damaged.The procedure was completed without the mapping catheter and there were no adverse consequences to the patient.
 
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.The electrodes were displaced and the pellethane tubing and splines were corrugated and torn.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the withdrawal difficulty, torn pellethane tubing, and displaced electrodes remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12852079
MDR Text Key281132315
Report Number3005334138-2021-00706
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD-AVHD-DF16
Device Lot Number8041962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received11/22/2021
Supplement Dates Manufacturer Received02/08/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexFemale
Patient Weight70 KG
-
-