• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT OBTURATOR W LASR; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT OBTURATOR W LASR; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Model Number 810081L
Device Problem Material Frayed (1262)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Event Description
It was reported that a patient underwent a gynecological procedure on an unknown date in 2021 and the mesh was used.It was reported that one side of the mesh couldn't be moved, and it frayed.There were no adverse patient consequence reported.Additional information was requested.
 
Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.-any patient consequences? -please provide procedure date -will the product be returned? if so, please provide tracking information.
 
Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 01/26/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 - device not returned.Additional information was requested and the following was obtained: any patient consequences? no.Please provide procedure date - unk.Will the product be returned? if so, please provide tracking information.No device received yet.
 
Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 03/29/2022.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 04/27/2022 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information:  d9, h3, h6 h3 evaluation: device was received for evaluation.The received device was manipulated and ex-planted as original packaging and several components were missing (box, ifu, blister, needles and mesh).Only one part of blue mesh and lid were returned.Some organic matter is visible in the returned sample.The remaining blue mesh was damaged and small parts were fraying.The defect on device seen during the product evaluation is aligned with the defect described in the event description.However, the defect identified is not linked to a manufacturing issue.Events of this type are trended regularly, therefore this complaint is being closed to trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT OBTURATOR W LASR
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20
neuchatel 2000
SZ   2000
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key12856544
MDR Text Key285684334
Report Number2210968-2021-11891
Device Sequence Number1
Product Code OTN
UDI-Device Identifier10705031062306
UDI-Public10705031062306
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K033568
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number810081L
Device Catalogue Number810081L
Device Lot Number3939104
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/29/2021
Initial Date FDA Received11/22/2021
Supplement Dates Manufacturer Received01/26/2022
03/02/2022
03/30/2022
Supplement Dates FDA Received01/26/2022
03/29/2022
04/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-