Catalog Number 2C1009K |
Device Problems
Crack (1135); Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that on the second day after starting a chemotherapy treatment on a patient with a large volume english-chinese infusor, the nurse found five (5) ml of chemotherapy drugs between the inner wall of the infusor and the bladder.After inspection, there was a crack in the fill port, and the bladder wall had a small hole for slow seeped liquid.As a result, the dose of the un-injected chemotherapy drug was calculated and the patient was given a supplementary infusion.The issue was identified during a patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|