• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Failure to Power Up (1476); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
As reported, the device button (start/stop button) is non functional, it cannot be switch on.The issue found at preparation for use.There is no patient involvement associated on this reported event.No user injury reported.
 
Manufacturer Narrative
The subject device was received and evaluated at rrc (regional repair center) olympus prerov.Device inspection found the front panel does not work properly, traces of liquid in connector of front panel observed.The front panel, blue bar and flat cable have to be replaced.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide a correction for g2 and additional information based on the legal manufacturer's final investigation.G2 - checked 'other' to add the country czech republic.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the nonfunctional button occurred from a damaged flat cable.The flat cable was rusty and could no function properly.The cause of the rusted flat cable is likely due to liquid penetration inside the device.A definitive root cause of the liquid present within the device could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12866596
MDR Text Key284447065
Report Number8010047-2021-14984
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/02/2021
Initial Date FDA Received11/23/2021
Supplement Dates Manufacturer Received12/16/2021
Supplement Dates FDA Received12/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/11/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-