• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH MATRIXMANDIBLE BENDING SCREW; BONE PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH MATRIXMANDIBLE BENDING SCREW; BONE PLATE Back to Search Results
Catalog Number 03.503.080
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2021
Event Type  malfunction  
Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021 after bending the plate, the bending screws should be removed but it was not possible due to the holes of the plate were deformed after bending process.The removal of the bending screws was only possible by drilling.Procedure was completed successfully with thirty(30) minutes surgical delay.Patient status is fine.This report is for one (1) matrixmandible bending screw.This is report 4 of 4 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MATRIXMANDIBLE BENDING SCREW
Type of Device
BONE PLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12869491
MDR Text Key285221619
Report Number8030965-2021-09758
Device Sequence Number1
Product Code JEY
UDI-Device Identifier07611819979840
UDI-Public(01)07611819979840
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K063790
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.503.080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/29/2021
Initial Date FDA Received11/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
BENDING INSERT HOLDER FOR 03.503.080; MATRIXMANDIBLE BENDING SCREW; UNK - PLATES: MATRIXMANDIBLE
Patient Weight70 KG
-
-