STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN 52MM SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
Catalog Number UNK_JR |
Device Problem
Malposition of Device (2616)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 11/01/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
|
|
Event Description
|
It was reported the patient's right hip was revised due to notching of the femoral stem due to impingement with the mdm liner.It was reported that this was likely the result of the shell being in too vertical of a position.The shell, mdm liner, adm/ mdm poly, and femoral head were revised.Rep confirmed there are no allegations against the poly insert or femoral head.
|
|