MAUDE Adverse Event Report: COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980X1JADIJJ
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2021
Event Type  malfunction  
Manufacturer Narrative
Covidien/medtronic has not received the device/component from the customer for evaluation.
 
Event Description
It was reported that during use on patient this 980 ventilator illuminated green alarm banner on gui (graphical user interface) with no audio sound generated.It was also mentioned that the device touch panel sound and dial knob sound was not working.Although patient intervention details are unknown, there was no harm to the patient.Upon reviewing logs it was confirmed that no audio alarm was generated at the time of reported event.
 
Manufacturer Narrative
Additional information section d9 additional information: h9 fda z-0966-2022 correction to initial report due evaluation: section h6 evaluation code method, result, conclusion and component codes.H3 device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that during use on patient this 980 venti lator illuminated green alarm banner on gui (graphical user interface) with no audio sound generated.It was also mentioned that the device touch panel sound and dial knob sound was not working.The device was available for evaluation.The service personnel (sp) inspected the ventilator and could not duplicate the reported issue.The sp updated software installation rev ap.Ventilator passed all test and calibrations per manufacturer specifications at the time of service.Further investigation into the reported event determined a plausibility of no audible alarm likely caused due to design issue.There is an existing previous internal investigation regarding this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality s pecifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during use on patient this 980 ventilator illuminated green alarm banner on gui (graphical user interface) with no audio sound generated.It was also mentioned that the device touch panel sound and dial knob sound was not working.Although patient intervention details are unknown, there was no harm to the patient.Upon reviewing logs it was confirmed that no audio alarm was generated at the time of reported event.
 
Manufacturer Narrative
In a previous supplemental report, the capa number and reason for the remedial action failed to be included.This report is being submitted in order to provide that information as a correction in field h10.During a regular internal review of complaints, it was identified that some medwatch reports associated with an fca were submitted without a correction/removal number in field h9 of 3500a form.Medtronic have initiated capa pr 576617 to address this gap.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
kelly adams
2101 faraday ave
carlsbad, CA 92008
7606035046
MDR Report Key12872280
MDR Text Key283943309
Report Number8020893-2021-00199
Device Sequence Number1
Product Code CBK
UDI-Device Identifier10884521171558
UDI-Public10884521171558
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number980X1JADIJJ
Device Catalogue Number980X1JADIJJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2021
Initial Date FDA Received11/24/2021
Supplement Dates Manufacturer Received11/11/2021
11/11/2021
Supplement Dates FDA Received03/30/2023
01/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0966-2022
Patient Sequence Number1
-
-