In the several days after, the water filter of the subject device was replaced by the field service engineer of olympus and it was found that the water filter had been clogged.Also it was found there was sufficient cleaning because the user checked the disinfectant solution to be effective with the test strip every time.Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.Based on the report of user and the field service engineer of olympus, omsc confirmed the user facility had not replaced the water filter of the subject device for over a year and found the water filter was clogged by the user handling.The water filter need to be replaced at least once a month.If additional information is received, this report will be supplemented.
|