B. BRAUN SURGICAL, S.A. DAFILON BLUE 4/0 (1.5) 45CM DS16; OTHER SUTURE Back to Search Results
Model Number C0932132
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available a follow-up report will be submitted.
 
Event Description
It was reported an issue with dafilon suture.The client reported that the needle was detached from the thread while the last knot was performing during gingival closure.The suture technique was simple interrupted stitches.There was no patient injury and no additional information has been provided.
 
Manufacturer Narrative
Analysis and results: there are no previous complaints of this code-batch.We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have received an open sample with the needle detached from the thread and the thread is not wound on the pack.However, without any closed sample we cannot carry out an analysis in order to take a decision.We have checked the thread and we have noticed that there are marks of a needle holder or other surgical instrument near the needle attachment area.The thread has been damaged during handling and it broke most probably due to a wrong handling not according to the instructions for use of the product.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.As stated in the instructions for use of the product, 'when working with suture materials great care should be taken to ensure that the use of surgical instruments, such as tweezers and needle holders, does not cause the material to be damaged by being pinched or kinked.' final conclusion: in spite of receiving a defective sample, without closed samples a suitable analysis cannot be performed.The case is considered not confirmed by evidence of the sample received.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 4/0 (1.5) 45CM DS16
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key12880700
MDR Text Key281323294
Report Number3003639970-2021-00624
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeHU
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0932132
Device Catalogue NumberC0932132
Device Lot Number621132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received11/29/2021
Supplement Dates Manufacturer Received02/16/2022
Supplement Dates FDA Received03/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-