The reported event was confirmed as manufacturing related.1 sample were confirmed to exhibit the reported failure.The reported failure is considered out of specification as the reported failure was reproduced.The product was not used for patient treatment.Visual evaluation of the returned sample noted one opened (without original packaging), three-way temp sensing silicone foley.Visual inspection of the sample noted attempt to infuse the balloon with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and was unable to advance the solution through the inflation lumen.Noted blockage in the inflation arm.This is out of specification which states, "check for defects: excessive material (over fill), bubbles/voids, splits and/or blemishes".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.A labeling review was not performed because labelling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|